Daily capacity up to 500 L in 8 hours



Wastewater treatment
Laboratory wastewater treatment system up to 500 L/day
BK-SFS500
Automatic system for treating up to 500 L of laboratory wastewater in 8 hours, with neutralisation, oxidation, disinfection, multi-stage filtration and advanced treatment modules.
Automatic pH, level and reagent control
Aeration, microelectrolysis and optional photocatalysis
Technical data
Reference configuration
The final configuration depends on daily flow, effluent composition, contaminant load, reagents and discharge requirements.
| Treated water volume | 500 L/day (8 hours) |
|---|---|
| pH measurement range | 0 to 14 |
| pH adjustment range | 6 to 9 |
| pH resolution | 0.01 pH |
| pH accuracy | ±0.02 pH |
| Treatment process | Neutralisation precipitation + combined disinfection by ozone/oxidising agent + UV sterilisation + multi-stage filtration + aeration + optional photocatalysis + microelectrolysis |
| System flushing | Positive and negative flushing |
| Water inlet | Suction or gravity inlet |
| Antifreeze protection | Automatic heating below 5 °C |
| Liquid-level monitoring | Multi-channel detection, reagent-shortage alarm and automatic treatment |
| Electrical supply | AC 220 V, 50/60 Hz; AC 110 V, 50/60 Hz optional |
| Package dimensions | Machine: 1500 × 710 × 1415 mm; accessories: 515 × 515 × 1315 mm |
| Gross weight | 220 kg |
Project and supply
Configured for your process
We confirm the effluent source and composition, flow rate, pH, critical contaminants, discharge point and monitoring options before the technical proposal.
- Effluent and flow-rate characterisation
- Definition of treatment stages and reagents
- Delivery across Europe, North America and South America